Spectrophotomètre Portable MYIRO-1
Nos instruments sont intégrés dans de nombreuses solutions logicielles.
Vous pouvez vérifier ici si votre logiciel préféré supporte nos instruments.
Si notre appareil n'est pas connecté à votre solution, demandez à votre fournisseur de nous contacter.
Cette liste est une première information. Veuillez vérifier auprès du fournisseur du logiciel si la fonctionnalité dont vous avez besoin est prise en charge.
Illumination/viewing system | 45°a:0° (annular illumination)*1 |
Spectral separation device | Concave grating |
Wavelength range | Spectral reflectance: 380 to 730 nm Spectral irradiance: 360 to 730 nm |
Wavelength pitch | Reflectance: 10 nm |
Half bandwidth | appr. 10 nm |
Measurement area | ø 3.5 mm |
Light source | LED |
Measurement range | Reflectance: 0 to 150% / Density: 0.0 D to 2.5D |
Repeatability | Colorimetric: Within 𝞼∆E00 0.05 (When white calibration plate is measured 30 times in 10-second intervals after white calibration has been performed) |
Inter-instrument agreement | Within ∆E00 0.3 (Average of 12 BCRA Series II color tiles compared to values measured with a master body under manufacturer's standard conditions) |
Measurement time (single-point) | Appr. 1s |
Measurement conditions *2 | M0, M1, M2, User-defined illuminant |
Observers | 2° Standard Observer, 10° Standard Observer |
Status indicator | LED to indicate instruments status |
Interface | Wireless LAN (802.11 b/g/n); USB 2.0 |
Scanning measurements | Scanning of color charts under all measurement conditions in one single scan |
Power | USB bus power; Rechargeable internal battery |
Dimensions (WxDxH) | 73 mm x 171 mm x 71 mm |
Operating temperature/humidity range | 10° to 35°C, 30% to 85% relative humidity with no condensation |
Storage temperature/humidity range | 0° to 45°C, 0% to 85% relative humidity with no condensation |
*1 | Illumination for wavelengths under 400 nm is unidirectional |
*2 | M0, M1, M2: Illumination conditions defined in ISO 13655 |